Autoclave
Autoclave (Industrial Use, Large Capacity) (YYK500)
Product Summary
Used to remove residual air bubbles after affixing polarizer in LED production.
Features of Autoclave (Industrial Use, Large Capacity) (YYK500)
Features
No temperature overshoot, precision temperature uniformity available
Features
4 step working procedures:
(1) Preheat: temperature rising, no pressurizing
(2) Pressurizing: holding temperature, pressurizing
(3) Deaeration: holding temperature, deaerating
(4) End: temperature cooling, pressure dropping
Features
Adjustable air suction and exhaust speed
Features
Customized chamber dimensions
Safety features
Door open / close detection, door lock / unlock detection, higher pressure alarm, air inlet pressure detection, safety valve, independent overheat protector, electric leakage breaker
Specifications
Specifications
| Item Code | YYK500 |
|---|---|
| Model | YYK500 |
| Method | Heating + pressurizing |
| Specifications | Class-1 pressure container (AQSIQ pressure container verification) |
| Operating temp. range | Room temp. +10~70°C |
| Operating pressure range | 0.101~0.9MPa |
| Temp. distribution accuracy | ±3°C (at 50°C) |
| Max. temp. reaching time | Within 15min (adjustable) |
| Max. pressure reaching time | Within 20min (adjustable) |
| Internal dimension (effective) | ø500mm×850Lmm |
| Material | SUS304 stainless steel, internal polishing |
| Max. operating pressure | 0.9MPa |
| Hydraulic test pressure | 1.35MPa |
| Medium | Dry air (pressure: working pressure +0.05MPa or higher) |
| Opening / closing system | Manual clutch easy to operate |
| Pressurizing system | Controlled by pressure controller |
| Heating system | PID control |
| Stirring system | Stirred by centrifugal fan (Water-Cooling is not required for shaft seal, free-maintenance) |
| Control system | PLC control |
| Pressure gauge | Pressure range: 0 to 1.0MPa, accuracy: ±1% (with upper limit alarm contact) |
| Temp. controller | Digital setting and display, PID control |
| Pressure controller | Digital setting and display, ON/OFF control |
| Working timer | Time range: 99 hr 59min, Digital setting and display, |
| Temp. sensor output | 5 groups of K thermocouple output terminals |
| Internal dimensions(φ mm) | φ500mm |
| External dimensions(W×D×H mm) | 1000×1656×1546mm |
| Remarks(dimensions) | Capacity 850 L |
| Power Source | AC 220 3 phase V |
| Weight | 700kg |