Clean Oven - Yamato Scientific

Clean Oven

Clean Oven (DE411)

Product Summary Clean oven suitable for temperature test in dust-free environment. Features of Clean Oven (DE411) Features Adoption of anti-fouling caster (preventing wheel contamination during transportation) Features Improvement of visibility of HEPA filter replacement timing by three color indication. Features Improvement of safety by phase-reversal relay, detecting of incorrect wiring of power supply when installing. Features Cable diameter changed from 30mm I.D to 33mm I.D when compare withconventional equipments. Features More options when compare with conventional equipment.

Specifications

Specifications

Item Code212610
ModelDE411
Temp. distribution accuracy±2.5°C (at 260°C) JTM K05
Max. temp. reaching timeApprox. 70 min. (at 260°C)
HeaterStainless pipe heater
Cable hole33mm I.D. ×1
Timer1 min. to 99 Hrs. 59 min
Operation functionFixed temp., Auto-start, Auto-stop, Quick auto stop, program (Max.99 steps, 99 patterns, repeat)
Circulation methodForced air circulation
Operating temperature rangeRoom temperature +30~260°C
Temp. adjustment accuracy±0.3°C (at 260°C) JTM K05
Temp. fluctuation1.0°C (at 260°C) JIS
Temp. gradient±10°C(at 260°C) JIS
Clean levelClass 100 (when the temperature is stable)
InteriorStainless steel
Exterior materialCold rolled steel plate with melamine resin baking finish
Fan typeSirocco fan, Condenser type motor
Differential pressure meterAnalog type (0~300 Pa)
Exhaust port61mm O.D.×1
FilterHeat resistant HEPA filter (dust collect : more than 99.97% at 0.3μm )
CasterWheel Dia. 50mm anti-fouling caster
Temp. controllerPID control by micro processor
Temp. setting methodDigital setting with UP/DOWN key
Temp display methodDigital display
Other displayOperation monitor (operation condition graphic display by LED patterns)
Additional functionsCalendar timer (max. 24 Hrs.), Calibration off-set function, Integration time (max. 65535Hrs.), Display the amount of power consumption/CO2 discharge/Heater operation amount, Power failure recovery mode, User setting information save and recall
Heater circuit controlTriac zero-cross control
Temp. sensorK-thermocouple (double sensor)
Safety countermeasuresSelf diagnosis functions (Sensor, Fan, Heater, Relay, Triac, Automatic overheat prevention), Independent overheat prevention, Key lock function, Electric leakage breaker, Door switch, Reverse phase protection
Shelf support step number12 steps
Accessories Shelf plateStainless wire steel, 2 pcs
Accessories Shelf support4 pcs
Internal dimensions(W×D×H mm)450×450×450mm
External dimensions(W×D×H mm)700×1025×1570mm
Remarks(dimensions)91L
Power SourceAC220/380V
Three phase
Weight200kg