Inert Oven (DN411I )
Inert Oven (DN411I )
Product Summary
Suitable for Low Oxidation Environment
Features of Inert Oven (DN411I )
Suitable for curing process in a low oxygen atmosphere.
Suitable for heat insulation test and curing process up to 360°C.
Feature
Simple operation by interactive key input
Features
Operation monitor visualizes controller status, temp and temp. changing.
Features
Incorporates with maximum 99 steps, 99 patterns program controller with repeat function.
Features
Loaded with total operation hours timer.
Features
N2 gas flow amount controllable.
Specifications
| Item Code | 212819 |
|---|---|
| Model | DNE411 |
| Operating temp. range | Room temp. +20 to 210°C |
| Temp. adjustment accuracy | ±0.5°C (at 210°C) |
| Temp. distribution accuracy | ±2.0°C (at 210°C) |
| Max. temp. reaching time | Approx. 60 min. |
| Interior/Exterior material | Stainless steel /Electro-galvanized steel sheet with epoxy and melamine resin baking finish |
| Operation function | Fixed temperature operation, Program operation, Auto start, Auto-stop, Quick automatic stop |
| Program mode | Program operation; Maximum 32 steps, Repeat function |
| Additional functions | Timer Function (Actual Timer within 24hrs.), Total operation hours timer (to 65535hrs.), Calendar time ( 24 hours), |
| Shelf plate with standard load | Approx. 15kg/piece |
| Heat insulating material | Glass wool |
| Heater | Stainless pipe heater |
| Heater | 1.1kW |
| Fan Type | Scirocco fan |
| Motor | Condenser type |
| Cable hole/Air in-take hole | 33mm I.D. (the right side) 1pc. each |
| Temp. controller | PlD control by microprocessor |
| Temp. setting method | Digital setting by UP/DOWN key |
| Temp display method | Digital Display by Green LED |
| Timer/Min.division | 1 min. to 99 Hrs. 59 min. /1 min. |
| Additional functions | Calibration offset, Power discharge, CO2 discharge, Power failure return mode, User configuration information |
| Heater circuit control | Triac zero-cross control |
| Safety countermeasures | Self diagnosis functions (Temp. sensor abnormal, Heater disconnection, SSR- short, Main relay abnormal, Automatic overheating prevention), |
| Safety countermeasures | Key lock function, Independent overheat prevention, Electric leakage breaker with over current protection |
| Shelf support step number | 11 steps |
| Shelf support pitch | 30 mm |
| Accessories Shelf plate | Stainless steel, 2 pcs. |
| Accessories Shelf support | 4 pcs. (1 pair – 2 pcs. is fixed) |
| Method | Forced air circulation |
| Internal dimensions(W×D×H mm) | 450×450×450mm |
| External dimensions(W×D×H mm) | 580×645×860mm |
| Remarks(dimensions) | 90L |
| Power Source | AC220V 6A Single-phase |
| Weight | 63kg |