DESCRIPTION

Features

SPECIFICATION

ITEMSPECIFICATIONS
Titration methodVolumetric Karl Fischer reagent titration method
Titration control methodIntermittent addition of titrant while approaching the end point. (1) Control using a waiting interval timer of 1 – 99sec (INTERVAL). Titration stops when there is no change in the electrode potential during this interval. (2) Control by titration timer. (T-TIMER: 1 – 99min.)
End point detectionDC polarization potential difference detection
Buret capacity10mL Reproducibility ±0.01 mL (for full discharge, standard deviation for n=5)
Liquid contact materialPTFE, ceramic and glass
Number of burets1
Display screen2 lines, 16 digit LCD with backlighting
Display unitsmL (min. display is 0.001mL), mg H2O, %, or ppm
Measuring range100 ppm – 100%
Minimum volume of titrant addition0.005mL or 0.01mL
Calculation functionConcentration, factor standardization, statistics, recalculation
Titration cell capacity150mL
Minimum volume of solventApprox. 30mL
External input/outputTwo RS-232C ports Port 1 (HOST) for PC connection. Port2 (BALANCE) for Balance connection.
PrintersOptional thermal and impact printer – please inquire
Power supply / Power consumptionAC100 – 240V, 50/60 Hz, 100 VA
Dimensions310 (W) x 270 (D) x 300 (H) mm
WeightApprox. 6kg
Safety approvalCE approved